Is Your AI Software a Medical Device? A CDSCO Guide for Hospitals (2026)
When AI software is a medical device in India: CDSCO's July 2026 software guidance, Class A–D matrix, licensing, AI/ML rules and a hospital buyer checklist.

Key takeaways
- CDSCO finalised its Guidance Document on Medical Device Software (CDSCO/MD/GD/MDSW/01/2026) on 21 July 2026. It explains how the Medical Devices Rules, 2017 already apply to software; it is not a new law.
- Whether software is a device depends on the vendor's intended-use statement, not on whether it uses AI.
- Standalone software is classed A–D using a matrix of the healthcare situation (critical, serious, non-serious) and the significance of the output (treat or diagnose, drive, or inform clinical management).
- HIS, LIS and image-management systems used only for admin, storage or sharing are out of scope — unless they add medical functions such as image analysis or real-time monitoring.
- AI-based devices must disclose training, validation and test data, show performance on Indian sub-populations, and monitor for drift after launch.
Short answer: In India, AI software is a medical device only when its intended purpose is medical — diagnosing, treating, monitoring, or driving or informing clinical management. CDSCO's Medical Device Software guidance (issued 21 July 2026) classes such software from A to D by how serious the condition is and how much the output drives the decision. Admin AI such as CRM, scheduling, call handling, training and operational analytics is out of scope, but still falls under the DPDP Act.
Disclosure: Surgy Health builds AI software for hospitals. This guide explains the rules as they apply to any vendor, including us. It is a practitioner's summary, not legal or regulatory advice.
What changed in 2026
CDSCO released a draft guidance on medical device software in October 2025 and issued the final Guidance Document on Medical Device Software on 21 July 2026. The 62-page document covers standalone software, software embedded in or driving hardware, mobile and cloud software, and AI- and machine-learning-based software.
CDSCO states that the guidance "reflects current practices under the MDR-2017 and should not be misconstrued as a new regulatory control" (section 2.0). For hospitals, the practical change is clarity: there is now a published test for when the software you buy is a regulated device, and a published list of what its maker must disclose.
What counts as medical device software?
Software is medical device software (MDSW) when its maker intends it for a medical purpose — diagnosis, prevention, monitoring, treatment or alleviation of disease, or supporting clinical decisions. The guidance covers two kinds:
- Software as a medical device (SaMD): standalone software that performs the medical function itself, such as computer-aided detection on X-rays or ECG analysis, or AI that triages scans or screens for cancer.
- Software in or driving a medical device (SiMD): software that controls hardware, such as an insulin pump's control software, or an app that controls a device over Bluetooth or Wi-Fi.
The same algorithm can be in or out of scope depending on what the vendor claims it is for. A model that summarises a consultation for the record is documentation. The same model, marketed to suggest diagnoses, is informing clinical management.
What is out of scope
Section 5.2 of the guidance lists software that is not medical device software:
- Hospital or clinical information systems (HIS/CIS) used only for admission, scheduling, billing, communication or storing patient information
- Laboratory information systems and image-management systems that only manage, store or share data
- ERP and quality-management software
- General communication tools such as email and video calling
- General wellness software that makes no reference to diseases or conditions
The caveat that matters for hospitals: an HIS, LIS or image-management system with added medical functions — for example, image analysis as a diagnostic aid or real-time patient monitoring — may be a medical device. Check each module, not just the product name.
Out of scope does not mean unregulated. Admin AI that processes patient data is covered by India's Digital Personal Data Protection Act and Rules, and NABH's information management standards still apply.
How CDSCO classifies medical device software
Software that drives or influences a hardware device takes the same risk class as that hardware (section 7.1). Standalone software is classified with the matrix below (Table 2 of the guidance). Where more than one rule applies, the stricter class applies. If standalone software is meant for non-clinical users in a serious situation without specialist support, it may be treated as critical.
| Healthcare situation | Treats or diagnoses | Drives clinical management | Informs clinical management |
|---|---|---|---|
| Critical | Class D | Class C | Class B |
| Serious | Class C | Class B | Class A |
| Non-serious | Class B | Class A | Class A |
The Central Licensing Authority confirms the class. CDSCO publishes a classification list on its MD Online portal; if a product type isn't listed, the applicant can ask CDSCO to classify it.
Hospital AI examples on the matrix
These examples are illustrative. The actual class depends on the vendor's intended-use statement and CDSCO's decision.
| Situation | Treats or diagnoses | Drives clinical management | Informs clinical management |
|---|---|---|---|
| Critical | D — AI that diagnoses stroke on CT | C — ICU deterioration score that triggers an escalation protocol | B — sepsis risk flag shown to the ICU team |
| Serious | C — diabetic retinopathy screening that outputs a diagnosis | B — insulin dose calculator | A — drug-interaction alert in e-prescribing |
| Non-serious | B — AI that diagnoses a skin condition from a photo | A — physiotherapy exercise progression tool | A — symptom information display |
Where common hospital AI tools usually land
| AI tool | Usually | What tips it into scope |
|---|---|---|
| Hospital CRM, patient follow-up, WhatsApp messaging | Out of scope | Clinical triage or treatment advice to patients |
| AI voice agent for booking and reminders | Out of scope | Symptom assessment that routes care |
| Staff training and LMS | Out of scope | — |
| Operational analytics and KPI dashboards | Out of scope | Patient-level risk predictions used for care |
| AI medical scribe (transcribe, structure, summarise) | Usually out of scope | Suggesting diagnoses or treatment, or flagging clinical risks |
| Clinical decision support alerts | In scope, often Class A–B | Higher class if the situation is critical or the output drives care |
| Imaging AI (detection, triage, diagnosis) | In scope, often Class B–D | — |
| Early-warning scores (sepsis, deterioration) | In scope, often Class B–C | — |
Who licenses what
Table 5 of the guidance sets the licensing authority by activity and class:
| Activity | Class A | Class B | Class C | Class D |
|---|---|---|---|---|
| Manufacturing licence | State Licensing Authority | State Licensing Authority | CDSCO (Central Licensing Authority) | CDSCO (Central Licensing Authority) |
| Import licence | CDSCO | CDSCO | CDSCO | CDSCO |
| Test licence | CDSCO | CDSCO | CDSCO | CDSCO |
| Sale and distribution | State Licensing Authority | State Licensing Authority | State Licensing Authority | State Licensing Authority |
The guidance notes that Class A devices that are non-sterile and non-measuring are exempt from the licensing requirements; their makers should still check registration obligations on the MD Online portal. Commercial applications go through CDSCO's MD Online system; test licences for manufacture (Form MD-12/MD-13) and import (Form MD-16/MD-17) go through the National Single Window System.
What CDSCO expects from AI-based medical devices
For AI and machine-learning software, the guidance asks makers to:
- Disclose the data. Describe the composition of training, validation and test datasets, and state whether data is real-world, from public databases or literature, or synthetic (section 12.4.2).
- Prove it works in India. Show performance and robustness across sub-populations relevant to India; models trained or validated outside India need justification and may need Indian validation.
- Control change. Describe whether the model updates itself or is updated under the maker's control, and where applicable submit an Algorithm Change Protocol covering data management, retraining, re-validation and rollback.
- Watch for drift. Keep documented procedures to monitor performance after deployment, including algorithm drift and degradation (section 9.0).
- Secure it. Encryption in transit and at rest, access control, audit trails, threat modelling at design stage and safe behaviour during incidents; standards referenced include IEC 81001-5-1 and ISO/IEC 27001.
- State the cloud. For software delivered as a service, state whether the cloud host is MeitY-empanelled (section 12.1).
- Keep evidence current. Clinical evidence can include literature, guidelines, clinical investigations and real-world data; post-approval changes must be notified.
Hospital buyer checklist
Before you buy any AI tool that touches clinical care, ask the vendor for:
- The intended-use statement, in writing, exactly as filed or planned.
- The CDSCO class and how it was decided (Table 2 cell, or hardware it drives).
- Registration or licence number, or a dated plan if the product isn't yet registered.
- Dataset summary: where training and test data came from and how Indian patients were represented.
- Indian validation results, ideally from a hospital like yours.
- Update policy: how model changes are made, tested and notified to you; the Algorithm Change Protocol if one exists.
- Drift-monitoring reports you will receive after go-live, and how often.
- Hosting: cloud provider, region and MeitY empanelment status.
- Field safety process: how you'll be told about a safety issue and what you must do.
- Your contract terms: who carries liability for errors, and audit rights.
Then run the tool in shadow mode on your own anonymised data. A CDSCO registration tells you the device is lawfully on the market; it doesn't tell you how accurate it will be in your wards.
Add these questions to your healthcare AI RFP, and record each approved tool in your AI governance policy register.
Selling or buying in the Gulf too?
Saudi Arabia (SFDA) and the UAE (MOHAP, plus Abu Dhabi DoH and Dubai DHA policies) regulate clinical AI separately. See AI in healthcare in the UAE and Saudi Arabia: rules for hospitals.
FAQ
Frequently asked questions
Is AI software a medical device in India?
Only when its intended purpose is medical: diagnosing, treating, monitoring, or driving or informing clinical management. AI used for admin work such as scheduling, billing, CRM, training or operational analytics is not a medical device under CDSCO's July 2026 guidance, though the DPDP Act still applies.
What is SaMD under CDSCO rules?
Software as a medical device is standalone software intended for a medical purpose that performs that purpose without being part of a hardware device, such as AI that detects findings on X-rays. CDSCO covers it under the Medical Devices Rules, 2017, as explained in its 2026 software guidance.
How does CDSCO classify medical device software?
Software that drives hardware takes the hardware's class. Standalone software uses a matrix: the healthcare situation (critical, serious, non-serious) against the output's significance (treat or diagnose, drive, or inform clinical management), giving Class A (lowest risk) to Class D (highest).
Is an HIS or EHR a medical device?
Not when it is used only for admission, scheduling, billing, communication or storing patient information. An HIS module that adds medical functions, such as diagnostic image analysis or real-time patient monitoring, may be a medical device and should be assessed on its own.
Is an AI medical scribe a medical device?
A scribe that transcribes, structures and summarises consultations for the record is usually documentation, not a medical device. If it suggests diagnoses or treatments, or flags clinical risks, that function may inform clinical management and should be classified using CDSCO's matrix.
Does Class A software need a licence?
CDSCO's guidance notes that Class A devices that are non-sterile and non-measuring are exempt from the licensing requirements. Makers should still check registration obligations on CDSCO's MD Online portal.
Who licenses software medical devices — CDSCO or the state?
For manufacturing, the State Licensing Authority licenses Class A and B, and CDSCO (the Central Licensing Authority) licenses Class C and D. CDSCO issues import and test licences for all classes; sale and distribution licences come from the state.
What must AI device makers disclose about training data?
The composition of training, validation and test datasets, whether the data is real-world, public or synthetic, and how the model performs across sub-populations relevant to India. Models trained outside India need justification and may need Indian validation.
Does CDSCO approval mean the AI will be accurate in our hospital?
No. Registration shows the product is lawfully on the market and its maker has met the evidence requirements. Accuracy in your wards depends on your patients, workflows and data, so test it on your anonymised data and monitor it after go-live.
Related reading
- Hospital AI buyer's guide
- DPDP Act for hospitals
- Hospital AI governance policy template
- Healthcare AI RFP template and scorecard
- AI medical scribe software for hospitals
Sources
- CDSCO, Guidance Document on Medical Device Software, Doc No. CDSCO/MD/GD/MDSW/01/2026 (issued 21 July 2026)
- Nishith Desai Associates, From code to compliance: decoding India's medical device software guidance (3 August 2026)
- Emergo by UL, India CDSCO finalizes guidance on medical device software (30 July 2026)
- Pure Global, India CDSCO medical device software guidance 2026 (9 August 2026)
- Ministry of Electronics and IT, Digital Personal Data Protection Rules, 2025
This guide summarises published guidance for hospital buyers. It is not legal or regulatory advice; confirm classification and licensing questions with CDSCO or qualified regulatory counsel.

Written by
Mohammed Jamil Nasir
Founder – Product & Technology, Surgy Health
14+ years in Product, Design & Tech · PGC AI/ML, IIT-Guwahati · Global MBA, SP Jain · BE-CSE
Mohammed Jamil Nasir leads product and technology at Surgy Innovation Labs Private Limited, building AI-first, seamless healthcare products for hospitals and healthcare networks across India and the GCC. He writes about healthcare AI, accreditation, and clinical operations.
Connect on LinkedInKeep reading
More on Healthcare Compliance
Hand-picked next reads that build on this article, so you can go deeper before you decide.
- Healthcare Compliance
DPDP Act for Hospitals: What You Actually Need to Know
No certificate to buy, no outsourcing shortcut. What India's DPDP Act and DPDP Rules 2025 require of hospitals, the 13 May 2027 deadline, penalties and a department-wise checklist.
14 min readRead article - Healthcare Compliance
AI in Healthcare in the UAE and Saudi Arabia: Rules for Hospitals (2026)
How AI in healthcare is regulated in the UAE and Saudi Arabia: data localisation, DoH and DHA policies, SFDA approval, PDPL transfers and a buyer checklist.
10 min readRead article - Healthcare AI
Healthcare AI RFP Template and Vendor Scorecard (Free Excel Kit)
A free healthcare AI RFP template for hospitals: 25 vendor questions, a 9-criterion weighted scorecard with scoring rubric, a worked example and an Excel kit.
11 min readRead article